Accu-flow:
Yes
Auto Sampler:
No
Calibration curves:
Up to 30 calibration curves. Automatic and Manual Calibration functionality.
Catalysts:
Yes
Communication Capabilities:
USB, Ethernet: LIMS Connectivity and Printer (optional)
Concentration:
In ppm: 0.05, 1, 5, 10, 50
Consumable/Utility Gas:
Vacuum
Dimensions: Depth:
21 in. / 54 cm
Dimensions: Height:
16.5 in. / 42 cm
Dimensions: Width:
15.8 in / 40 cm
Element:
Chlorine / 0.13 ppm
Measurement Time:
User selectable 10 - 999 s, 300 - 600 s typical
Method Name:
ASTM D7536 and D4929C
Options:
USB Printer (Receipt or Full Page), Consumables, Calibration Standards and Spare Parts.
Range:
0.13 ppm - 4%
Repeatability:
ppm - r
Xylene
0.05 - 0.15
1 - 0.17
5 - 0.31
10 - 0.50
Crude Oil
5 - 0.4
10 - 0.8
50 - 1.2
Water
5 - 0.6
10 - 1.0
50 - 1.5
Reproducibility:
ppm - R
Xylene
0.05 - 0.27
1 - 0.29
5 - 0.53
10 - 0.90
Crude Oil
5 - 0.7
10 - 1.4
50 - 2.0
Water
5 - 1.0
10 - 1.5
50 - 2.5
Sample Cup Type:
Standard XRF Cup
Sample Matrix:
Hydrocarbon/ water
Specific Technology:
MWDXRF
Type:
Benchtop
Weight:
75 lbs / 34 kg
What's included?:
Quick-Start Guide, Operations Manual, radiation survey, power cord, 6' drain tubing, spare fan filter, spare fuse, 100 sample cups, 100 etnom films, 30 etnom HV accucells, accucell inserts, 30 syringe, sample stand, spare primary window and 1 calibration set; 0-500 ppm Cl in Mineral Oil (6 x 10 mL). Printer optional.
XRF Film Type:
Etnom