Calibration curves:
Supports up to 30 calibration curves. Automatic and manual calibration functionality.
Catalysts:
No
Communication Capabilities:
USB, Ethernet; LIMS Connectivity
Consumable/Utility Gas:
Vacuum
Dimensions: Depth:
21 in / 54 cm
Dimensions: Height:
16.5 in / 42 cm
Dimensions: Width:
15.8 in / 40 cm
Element:
Sulfur/ 0.12 ppm
Measurement Time:
User selectable 10-999 s
Method Name:
ASTM D2622, ASTM D7039, ISO 20884, EN 16997
Options:
USB Printer (Receipt or Full Page), Consumables, Calibration Standards, Spare Parts
Range:
0.12 ppm - 10%
Repeatability:
ppm - r
2 - 0.3
5 - 0.5
8 - 0.6
15 - 0.8
100 - 2
500 - 5
Reproducibility:
ppm - R
2 - 0.7
5 - 0.8
8 - 1.0
15 - 1.4
100 - 4
500 - 10
Sample Cup Type:
Accucell
Sample Matrix:
Hydrocarbon
Specific Technology:
MWDXRF
Type:
Benchtop
Weight:
75 lbs / 34 kg
What's included?:
Quick-Start Guide, operations manual, radiation survey, power cord, 6' drain tubing, spare fan filter, spare fuse, 100 Accucell Mylar Sample Cups (490002-01), sample stand, spare primary window and one calibration set, D7039 and D2622 S in Mineral Oil calibration sets.
XRF Film Type:
Etnom