Calibration curves:
Supports up to 30 calibration curves. Automatic and manual calibration functionality.
Catalysts:
No
Communication Capabilities:
USB, Ethernet; LIMS Connectivity
Consumable/Utility Gas:
Vacuum
Dimensions: Depth:
21 in / 54 cm
Dimensions: Height:
16.5 in / 42 cm
Dimensions: Width:
15.8 in / 40 cm
Element:
Sulfur/ 0.18 ppm
Measurement Time:
User selectable 10-999 s
Method Name:
ASTM D2622, ASTM D7039, ISO 20884
Options:
USB Printer (Receipt or Full Page), Consumables, Calibration Standards, Spare Parts
Range:
0.18 ppm - 10%
Repeatability:
ppm - r
2 - 0.3
5 - 0.5
8 - 0.6
15 - 0.8
100 - 2
500 - 5
Reproducibility:
ppm - R
2 - 0.7
5 - 0.8
8 - 1.0
15 - 1.4
100 - 4
500 - 10
Sample Matrix:
Hydrocarbon
Specific Technology:
MWDXRF
Type:
Benchtop
Weight:
75 lbs / 34 kg
What's included?:
Quick-Start Guide, Operations Manual, radiation survey, power cord, 6' drain tubing, spare fan filter, spare fuse, 100 Accucell Mylar Sample Cups, sample stand, spare primary window and two calibration sets, D7039 and D2622 S in Mineral Oil calibration sets.
XRF Film Type:
Mylar