XOS - US
Get eNewsletter   1.518.880.1500 Service

Sindie 2622 Gen2 with Autosampler - Accu-cell (vacuum)

OBSOLETE ITEM

Sindie 2622 Gen2 with Autosampler - Accu-cell (vacuum)
Series: 402169-01M

OBSOLETE ITEM

Easy to use

User-friendly with touch screen interface. Just plug-it-in and measure, results with one touch: power is only utility.

Extremely low maintenance

No conversion gasses, heating elements, quartz tubes or columns.

Total sulfur analysis

From ultra low sulfur diesel and gasoline to heavy fuel oil and crudes, the analyzer delivers unprecedented precision and accuracy.

For use in multiple application areas

Refinery labs, pipeline terminals, additive plants and inspection laboratories.

Hassle-free sample prep

Utilizes Accucell pre-assembled and pre-vented sample cups for enhanced precision, extreme ease-of-use and enhanced productivity.

What's in the box?

Quick-Start Guide, Operations Manual, radiation survey, spare fan filter, printer with ink, printer cable, 1 roll printer paper, system fuse, 100 accucell cups, 100 mylar sample films, sample stand, spare primary window and 2 sets calibration standards; 1 for 7039 and 1 for 2622 (6 x 10 mL each).

Specifications

Auto Sampler: Yes
Calibration curves: 8 calibration curves. Automatic and Manual Calibration functionality.
Catalysts: No
Communication Capabilities: Printer, Ethernet: LIMS Connectivity (optional)
Consumable/Utility Gas: Vacuum
Dimensions: Depth: 19.7 in / 50.0 cm
Dimensions: Height: 13.4 in / 34.0 cm
Dimensions: Width: 14.6 in / 37.0 cm
Element: Sulfur / 0.4 ppm
Measurement Time: User selectable 30-900 s, 300 s typical
Method Name: ASTM D2622,
ASTM D7039,
ISO 20884
Options: LIMS Connectivity
Range: 0.4 ppm - 10 wt%
Repeatability: ppm - r
4 - 0.4
8 - 0.7
15 - 0.9
100 - 3
500 - 6
Reproducibility: ppm - R
4 - 1.0
8 - 1.2
15 - 1.7
100 - 6
500 - 12
Sample Cup Type: Accucell
Sample Matrix: Hydrocarbon
Specific Technology: MWDXRF
Type: Benchtop
Weight: 73 lbs / 32 kg
What's included?: Quick-Start Guide, Operations Manual, radiation survey, spare fan filter, printer with ink, printer cable, 1 roll printer paper, system fuse, 100 accucell cups, 100 mylar sample films, sample stand, spare primary window and 2 sets calibration standards; 1 for 7039 and 1 for 2622 (6 x 10 mL each).