Accu-flow:
No
Auto Sampler:
No
Calibration curves:
8 calibration curves. Automatic and Manual Calibration functionality.
Catalysts:
No
Communication Capabilities:
Printer, Ethernet: LIMS Connectivity (optional)
Concentration:
In ppm: 5, 10, 50 (crude oil)
Consumable/Utility Gas:
Vacuum
Dimensions: Depth:
19.7 in / 50.0 cm
Dimensions: Height:
13.4 in / 34.0 cm
Dimensions: Width:
14.6 in / 37.0 cm
Element:
Chlorine / 0.13 ppm
Measurement Time:
User selectable 30-900 s, 300-600 s typical
Method Name:
ASTM D7536, ASTM D4929
Options:
LIMS Connectivity
Range:
0.13 ppm - 3000 ppm
Repeatability:
ppm - r
Xylene
1 - 0.17
5 - 0.31
10 - 0.50
Crude Oil
5 - 0.4
10 - 0.8
50 - 1.2
Water
5 - 0.6
10 - 1.0
50 - 1.6
Reproducibility:
ppm - R
Xylene
1 - 0.29
5 - 0.53
10 - 0.90
Crude Oil
5 - 0.8
10 - 1.4
50 - 2.4
Water
5 - 1.0
10 - 1.5
50 - 3.2
Sample Cup Type:
Accucell
Sample Matrix:
Hydrocarbon
Specific Technology:
MWDXRF
Type:
Benchtop
Weight:
70 lbs / 32 kg
What's included?:
Quick-Start Guide, Operations Manual, radiation survey, power cord, 6' drain tubing, spare fan filter, printer, printer cable, printer ink, 1 roll of printer paper, spare fuse, 100 sample cups, 100 sample films, sample stand, spare primary window and 1 calibration set; 0-500 ppm Cl in Mineral Oil (6 x 10 mL).
XRF Film Type:
Etnom