Accu-flow:
No
Auto Sampler:
No
Calibration curves:
Up to 30 calibration curves. Automatic and Manual Calibration functionality.
Catalysts:
No
Communication Capabilities:
USB, Ethernet; LIMS Connectivity
Consumable/Utility Gas:
Internal Vacuum Pump
Dimensions: Depth:
21 in / 54 cm
Dimensions: Height:
16.5 in / 42 cm
Dimensions: Width:
15.8 in / 40 cm
Element:
Chlorine / 0.13 ppm
Measurement Time:
User selectable 10 - 999 s, 300 - 600 s typical
Method Name:
ASTM D7536 and D4929C
Options:
USB Printer (Receipt or Full Page), Consumables, Calibration Standards, Spare Parts.
Range:
0.13 ppm - 4%
Repeatability:
ppm - r
Xylene
0.05 - 0.15
1 - 0.17
5 - 0.31
10 - 0.50
Crude Oil
5 - 0.4
10 - 0.8
50 - 1.2
Water
5 - 0.6
10 - 1.0
50 - 1.5
Reproducibility:
ppm - R
Xylene
0.05 - 0.27
1 - 0.29
5 - 0.53
10 - 0.90
Crude Oil
5 - 0.7
10 - 1.4
50 - 2.0
Water
5 - 1.0
10 - 1.5
50 - 2.5
Sample Cup Type:
Standard XRF Cup
Sample Matrix:
Hydrocarbon/ water
Specific Technology:
MWDXRF
Type:
Benchtop
Weight:
75 lbs / 34 kg
What's included?:
Quick-Start Guide, Operations Manual, radiation survey, power cord, 6' drain tubing, spare fan filter, spare fuse, 100 sample cups, 100 sample films, sample stand, spare primary window and 1 calibration set; 0-500 ppm Cl in Mineral Oil (6 x 10 mL). Printer optional.
XRF Film Type:
Etnom