Calibration curves:
One
Catalysts:
No
Communication Capabilities:
(2) 4-20mA; Modbus and Ethernet (optional)
Consumable/Utility Gas:
Instrument air or nitrogen
Dimensions: Depth:
18 in / 45.7 cm
Dimensions: Height:
60 in / 152.4 cm
Dimensions: Width:
38 in / 96.5 cm
Element:
Chlorine / 0.2 ppmw in hydrocarbons, 0.6 ppmw in aqueous streams
Measurement Time:
User selectable, 15-300 s typical
Method Name:
ASTM D7536
Number of sample streams:
One; up to four (optional)
Options:
Sample Conditioning System
Range:
0.2 ppmw - 3000 ppmw
Sample Matrix:
Hyrocarbon/water
Specific Technology:
MWDXRF
Type:
On-Line
Weight:
300 lbs / 136 kg
What's included?:
Base system includes:
- Clora Online MWDXRF analyzer
- Rated for area classification CLASS I Div II Groups B, C, D T4A
- Stainless steel nameplate
- Signal output: Two (2) user configurable 4-20 mA result outputs; up to 4 alarm outputs (volt-free)
- Manual sample valve for lab verification
- Sample management system
Modular valves for introduction of calibration fluids
Pressure and flow regulators
- One (1) reference literature library includes:
- One (1) Installation manual
- One (1) User manual
- One (1) Maintenance manual
- One (1) CD copy containing all of the above
- Includes one (1) high viscosity dynamic window module
XRF Film Type:
Kapton