XOS - US
Get eNewsletter   1.518.880.1500 Service

Sindie Online

Sindie Online
Series: 400615-01
Unit Price Get Quote

Low maintenance

No heating elements, quartz tubing, or columns

Continuous monitoring with programmable response times

30 seconds: pipeline interface cuts
5 minutes: most demanding refinery processes

Outstanding linearity

One calibration for diesel and gasoline matrixes over full dynamic range

Direct and non-destructive

No consumables, no gasses, no high temperature processes

Increase productivity

Direct measurement without sample conversion - Analysis in ppm (wt)

What's in the box?

Base system Includes:
- SINDIE 6010 On-Line MWD XRF analyzer based on ASTM method D 7039
- Rated for area classification CLASS I Div 2 Groups B, C, D T4A
- Stainless steel nameplate
- Dynamic range 0.6 - 3000 ppm wt.
- Signal output: (2) user configurable 4-20 mA result outputs
- Manual sample valve for lab verification
- Sample Management System
Modular valves for introductions of calibration fluids
Pressure and flow regulartors/meters
-Five (5) 300cc SS sample cylinders for calibration fluids
- One (1) reference literature library includes:
One (1) Installation Manual
One (1) User Manual
One (1) Maintenance Manual
One (1) CD copy containing all of the above
-Includes one (1) Dynamic Window Module

Specifications

Calibration curves: 3 - 5 point linear calibration curve
Catalysts: No
Communication Capabilities: (2) 4-20mA; Modbus and Ethernet (optional)
Consumable/Utility Gas: Instrument air or nitrogen
Dimensions: Depth: 18 in / 45.7 cm
Dimensions: Height: 60 in / 152.4 cm
Dimensions: Width: 38 in / 96.5 cm
Element: Sulfur / 0.5 ppm
Measurement Time: User selectable, 15-300 s typical
Method Name: ASTM D7039
Number of sample streams: One; up to four (optional)
Options: Sample Conditioning System, High Range
Range: 0.5 ppmw - 3000 ppmw; Extended Range (XR) available
Repeatability: ppm - r
2 - 0.6
8 - 1.0
15 - 1.5
100 - 5
500 - 9
Sample Matrix: Hydrocarbon
Specific Technology: MWDXRF
Type: On-Line
Weight: 280 lbs / 127 kg
What's included?: Base system Includes:
- SINDIE 6010 On-Line MWD XRF analyzer based on ASTM method D 7039
- Rated for area classification CLASS I Div 2 Groups B, C, D T4A
- Stainless steel nameplate
- Dynamic range 0.6 - 3000 ppm wt.
- Signal output: (2) user configurable 4-20 mA result outputs
- Manual sample valve for lab verification
- Sample Management System
Modular valves for introductions of calibration fluids
Pressure and flow regulartors/meters
-Five (5) 300cc SS sample cylinders for calibration fluids
- One (1) reference literature library includes:
One (1) Installation Manual
One (1) User Manual
One (1) Maintenance Manual
One (1) CD copy containing all of the above
-Includes one (1) Dynamic Window Module
XRF Film Type: Kapton