The Evolution of Polycapillary Optics

Written by XOS Team | Jun 2, 2026 9:37:31 PM

 

The concept of polycapillary optics was introduced in the late 1980’s. During this time, XOS and a few other groups around the world conducted extensive research and development to significantly improve the optic manufacturing process. As a result, XOS became a leading global manufacturer of highperformance X-ray polycapillary optics and excitation systems. Today, polycapillary X-ray optics are used as key components in high-end X-ray analytical instruments and have dramatically enhanced system performance.

Background

A polycapillary X-ray optic consists of an array of tiny glass capillaries (up to a few million) formed into a well-defined shape. It collects a large solid angle of X-rays from a divergent X-ray source and redirects them by multiple external total reflections, to form either a focused beam (Figure 1) or a quasi-parallel beam (Figure 2).

Figure 1: Focusing X-ray Optic

A polycapillary focusing optic can focus the X-rays to a spot size as small as 5µm (at Rh Ka radiation, 20.2 keV). The X-ray flux density obtained at the focus of the optic can be up to five orders of magnitude higher than that obtained with a conventional pinhole collimator. This level of X-ray flux density improvement has been a gamechangernger for micro X-ray fluorescence.


Figure 1

The most important aspects of a polycapillary optic are the design and performance parameters:

  • Input focal distance (IFD): To define where the X-ray source is to be placed

  • Output focal distance (OFD): To define where the smallest beam size is obtained

  • Focal spot size: Energy dependent and normally characterized using the knife-edge method

  • Effective collection solid angle (ECSA): Directly proportional to the output beam intensity

  • Intensity gain: Comparison to a conventional pinhole collimator

  • Transmission efficiency: Ratio of the X-rays exiting the output end of the optic to those entering the optic at the input end

Figure 2: Collimating X-ray Optic

A polycapillary collimating optic, with the output beam diameter as large as 25mm, can generate a quasi-parallel X-ray beam for applications such as X-ray diffraction (XRD) and wavelength dispersive spectrometry (WDS). Depending on the X-ray energy, the collection solid angle of the optic can be up to 40˚ and the transmission efficiency can reach 40%.


Figure 2

Conclusion

The advances of the polycapillary X-ray optics have made it a must-have optical component in high-end X-ray analytical instruments for pioneering industrial applications like micro XRF. The optics have also been successfully used in synchrotron radiation facilities for research in micro X-ray analysis (XRF and X-ray absorption spectroscopy). Other unique applications include confocal XRF analysis and super-conducting EDS detectors.

Product Highlight:
Polycapillary Optics

Improve X-ray analysis performance and capability
A polycapillary optic captures a large solid angle of X-rays from an X-ray source and redirects them to a micron-sized focal spot or a highly collimated beam. The X-ray intensity achieved with such optics is a few orders of magnitude higher than that obtained with conventional pinhole collimators, contributing to the significantly improved X-ray analysis performance in detection sensitivity, spatial resolution, measurement speed, and precision. XOS optics are widely used in commercial instruments and customized X-ray analysis systems for various industrial and research applications in the fields such as microelectronics, semiconductor
manufacturing, pharma, and life sciences.